Influencia de correcciones IMAR por artefacto metálico en cálculo dosimétrico de Sistemas de Planificación Radioterápica Pinnacle

Ortiz Seidel, Mónica. (2020). Influencia de correcciones IMAR por artefacto metálico en cálculo dosimétrico de Sistemas de Planificación Radioterápica Pinnacle Master Thesis, Universidad Nacional de Educación a Distancia (España). Facultad de Ciencias

Ficheros (Some files may be inaccessible until you login with your e-spacio credentials)
Nombre Descripción Tipo MIME Size
Ortiz_Seidel_Monica_TCI.pdf Ortiz_Seidel_Monica_TCI.pdf application/pdf 5.12MB

Título Influencia de correcciones IMAR por artefacto metálico en cálculo dosimétrico de Sistemas de Planificación Radioterápica Pinnacle
Autor(es) Ortiz Seidel, Mónica
Resumen La terapia con radiaciones ionizantes, radioterapia, precisa para conseguir el objetivo de control tumoral una precisión en torno a un 3-4% en el cálculo de dosis. Los algoritmos actuales basan dicho cálculo en la información que aporta la imagen de tomografía computerizada (TC), tanto la de tipo morfológico, fundamental para la definición de contornos, como la relativa a los valores Hounsfield de cada pixel, que el sistema de planificación relacionará con la densidad electrónica relativa del medio, necesaria para el cálculo. La presencia en el cuerpo humano de prótesis metálicas produce en la imagen TC una distorsión de ésta conocida como artefacto metálico. El presente trabajo tiene como objetivo establecer si las correcciones IMAR a estos artefactos que permite realizar el tomógrafo Somaton Confidence (Siemens), redundan no sólo en una imagen mejorada que haga más fiable la labor de contorneo de los volúmenes de tratamiento y de riesgo, sino también en un cálculo más preciso de la dosis administrada. Para ello se estudian ejemplos representativos de las diversas situaciones: prótesis dentales, de cadera y expansores mamarios.
Abstract Therapy with ionizing radiation, radiotherapy, must have a precision of 3-4% in dose calculation in order to achieve the goal of tumoral control. Current algorithms are based on the information supplied by tomographic image, which is use morphologically for contour definition, as well as quantitatively as relative electron density of the medium, needed for calculation, is derived from Hounsfield units on the CT. The presence of metallic prosthesis on the human body produces distortions on the CT image known as metallic artifacts. This study aims to establish whether the IMAR corrections performed by CT Somaton Confidence (Siemens) to such artifacts leads, not only to a more reliable image to contour volumes of treatment and at risk, but also to an improved dose calculation. Some representative examples will be covered: dental implants, hip prostheses and temporary mammary expanders.
Notas adicionales Trabajo de Comienzo de Investigación. Máster Universitario en Física Médica. UNED
Materia(s) Física
Editor(es) Universidad Nacional de Educación a Distancia (España). Facultad de Ciencias
Director/Tutor Santa Marta Pastrana, Cristina
Fecha 2020-10-09
Formato application/pdf
Identificador bibliuned:master-Ciencias-FMed-Mortiz
http://e-spacio.uned.es/fez/view/bibliuned:master-Ciencias-FMed-Mortiz
Idioma spa
Versión de la publicación acceptedVersion
Nivel de acceso y licencia http://creativecommons.org/licenses/by-nc-nd/4.0
info:eu-repo/semantics/openAccess
Tipo de recurso master Thesis
Tipo de acceso Acceso abierto

 
Versiones
Versión Tipo de filtro
Contador de citas: Google Scholar Search Google Scholar
Estadísticas de acceso: 409 Visitas, 488 Descargas  -  Estadísticas en detalle
Creado: Tue, 27 Oct 2020, 00:08:51 CET