Control de calidad automático para los sistemas de Resonancia Magnética mediante el uso del Fantoma de Fluke 76-903

Rodríguez Peña, Francisco Javier. (2019). Control de calidad automático para los sistemas de Resonancia Magnética mediante el uso del Fantoma de Fluke 76-903 Master Thesis, Universidad Nacional de Educación a Distancia (España). Facultad de Ciencias

Ficheros (Some files may be inaccessible until you login with your e-spacio credentials)
Nombre Descripción Tipo MIME Size
Rodriguez_Pena_FcoJavier_TFM.pdf Rodriguez_Pena_FcoJavier_TFM.pdf application/pdf 2.62MB

Título Control de calidad automático para los sistemas de Resonancia Magnética mediante el uso del Fantoma de Fluke 76-903
Autor(es) Rodríguez Peña, Francisco Javier
Resumen Las medidas realizadas con fantoma nos permiten estudiar el comportamiento global del escáner de RM. Especialmente nos va a permitir comparar los resultados obtenidos en diferentes sistemas de RM, además de analizar la reproducibilidad de los resultados en un escáner determinado. Esto nos permite evaluar la caída de calidad de un escáner ya que se dispone de los datos de los controles de calidad a lo largo del tiempo, realizados de la misma manera y, por lo tanto, comparables. Estas características requieren consistencia en la adquisición de imágenes así como un protocolo de control de calidad. El objetivo de este trabajo ha sido desarrollar un programa que realice de forma automática el cálculo de determinados parámetros fundamentales para un control de calidad de un sistema de RM: relación señal/ruido, uniformidad de la imagen, imágenes residuales, resolución espacial y linealidad. Todos los procedimientos han sido comparados con el método manual utilizado en muchos departamentos de Física hospitalaria, obteniéndose resultados consistentes, generalmente bien correlacionados con los resultados obtenidos por los operadores.
Abstract Phantom measurements allow us for investigating the overall performance of an MRI scanner. Especially, it allow us to compare the results across different sites, in addition to the performance stability of a single scanner over time. This allows us to evaluate the fall in quality of a scanner since the data of the quality controls are available over time, carried out in the same way and, therefore, comparable. These characteristics require consistency in the acquisition of images as well as a quality control protocol. The objective of this work has been to develop a program that automatically calculates certain fundamental parameters for quality control of an MRI system: signal-to-noise ratio, image uniformity, residual images, spatial resolution and linearity. All the procedures have been compared with the manual method used in many Medical Physics departments, obtaining consistent results, generally well correlated with the results obtained by the operators.
Notas adicionales Trabajo Fin de Máster. Máster Universitario en Física Médica. UNED
Materia(s) Física
Editor(es) Universidad Nacional de Educación a Distancia (España). Facultad de Ciencias
Director/Tutor Santa Marta Pastrana, Cristina
Macías Verde, David
Fecha 2019-02
Formato application/pdf
Identificador bibliuned:master-Ciencias-FMed-Fjrodriguez
http://e-spacio.uned.es/fez/view/bibliuned:master-Ciencias-FMed-Fjrodriguez
Idioma spa
Versión de la publicación acceptedVersion
Nivel de acceso y licencia http://creativecommons.org/licenses/by-nc-nd/4.0
info:eu-repo/semantics/openAccess
Tipo de recurso master Thesis
Tipo de acceso Acceso abierto

 